Store
Global Offices

MFA-70Xi

Marine Fuel Analyzer: Cloud, Pour, Viscosity & Density

DFA

State-of-the-Art Solution

PAC Powered by Phase Technology’s MFA-70Xi analyzer is the world’s first and only instrument that is specifically designed to test four important properties of marine distillate fuels in one single unit —cloud point, pour point, viscosity and density. And, it does it all in less than 25 minutes!

Product Overview

Marine distillate fuels, or sometimes refer to as Marine Gas Oil (MGO), are required to have less than 0.5% of sulfur, a significant decrease from the previous 3.5% prior to 2020. It is critical to ensure these low sulfur MGO’s cold flow operability is suitable for the cold temperature environment it will experience throughout the ship’s journey.

Phase Technology’s MFA-70Xi analyzer is the world’s first and only instrument that is specifically designed to test four important properties of MGO’s in one single unit —cloud point, pour point, viscosity and density. And, it does it all in less than 25 minutes!

Simple and easy to use, the MFA-70Xi performs all four tests with the single push of a button. Sample is automatically loaded, without the need of a pipette. An optional 48-position autosampler offers greater throughput and improved automation if desired. The MFA-70Xi is completely self-cleaning; no solvent is required.

The MFA-70Xi was developed to meet the growing needs of the marine fuel industry as they test MGO distillate fuels. This includes DMX, DMA, DMZ, and DMB (also known as DFA, DFZ, DFB).

DFA

Key Features

1

Super fast testing speed

2

Advanced viscosity capability

3

Automatic sample input

4

Self cleaning – no solvent required.

DFA
5

No external chiller – Quiet, cool, self-contained thermoelectric cooler

6

No breakable glassware with internal capillary system

7

Superior precision with best measured repeatability and reproducibility

Specs

Test Methods & Precision

Standard Test Methods:
Cloud Point: ASTM D5773 (IP 446), ASTM D2500 (IP 219/ISO 3015) equivalent or better
Pour Point: ASTM D5949, ASTM D97 (IP 15/ISO 3016) equivalent or better
Kinematic Viscosity: ASTM D7945 @ 40°C, ASTM D445 (IP 71/ISO 3104) equivalent or better
Density: ASTM D7777, ASTM D4052 (IP 365/ISO 12185) @ 15°C correlation

Repeatability & Reproducibility:
Cloud Point: Repeatability:1.3 ºC,Reproducibility: 2.5 ºC
Pour Point: Repeatability:1.6 ºC @ 1 ºC,2.2 ºC @ 3 ºC,Reproducibility: 3.2 ºC @ 1 ºC,3.8 ºC @ 3 ºC
Kinematic Viscosity: Repeatability:0.5% @ 40°C,Reproducibility: 0.8% @ 40°C
Density: Repeatability:0.006081 (1000 – x) kg/m3,Reproducibility: 0.008545 (1000 – x) kg/m3

Bias: 0 (relative to ASTM manual methods)

Operation

Detection:
Diffusive Light Scattering (DLS) technology for freeze point
Horizontal capillary with motive force for viscosity

Cooling System:
Integrated Peltier device cooling system
Linear cooling profiles according to freeze point test method

System Cleaning:
Automatic flush cycle, no solvent required
Application Range

Cloud & Pour Point Temperature Range
Sample: – 80 to +70°C °C or °F (user selectable)
Test Resolution 0.1°C

Measurements

Sample Size:
0.15 mL required for test
minimum 20 mL including cleaning

Test Duration:
Cloud Point & Density combined: 5 to 10 minutes
Viscosity & Density combined: 10 minutes
Cloud, Pour, Viscosity & Density combined: 25 minutes

Required Operator Time: 0.25 minutes

Documentation

Printing:
Label printer: Detailed test report in °C or °F with date & time stamped, sample information, test mode
Full-size printer: Detailed detection curve tracking of fuel behaviors during the test; can be printed as part of test report.

Internal Memory: Storage up to 5000 test runs

QC-Functions: Automatic QC-sample handling and QC-Chart

Interface

Display:
Full-color, touch-sensitive, 15” high-resolution LCD screen with virtual keyboard
Languages: English, Chinese

Printer: USB interface for label printer or full-size HP/PCL printer

LIMS interface: Ethernet or RS232, Automatic LIMS transfer

Data Export: USB Memory Stick, Import into Excel

Password Protection: Different access levels for Operator, Manager, Service

Diagnostics: Automatic diagnostic functions

Accessories: Barcode Reader, label printer

Utility Requirements

Electrical: 90V to 260 VAC; 47-63Hz; 350W

Ambient Conditions: Temperature: max. 10° to 30°C; extremes not recommended

Dimensions: 54.6 cm (21.5”) deep x 33.7 cm (13.25”) wide x 44.5 cm (17.5”) high

Weight:
Unit: 24 kg (53 lb.)
Boxed: 28 kg (62 lb.)

Documentation

Title Description PDF
Brochure  MFA 70Xi Brochure Download
Application Notes Solutions to comply with IMO 2020 standards Download

Request a Demo or a Quote

img img img img check