State-of-the-Art Solution
OptiMVD is a mini viscometer and density meter designed for today’s modern laboratory that maximizes the power of automation to increase test productivity with significant repeatability and reproducibility improvement.
Product Overview
Fast and Reliable Viscosity Measurement
OptiMVD is a mini viscometer and density meter designed for today’s modern laboratory. Available in five different configurations, all models load the sample, perform viscosity and density measurement, and clean the system automatically. OptiMVD maximizes the power of automation to increase test productivity with significant repeatability and reproducibility improvement.
The dual viscometer/density meter model features two integrated, 24-position autosampler carousels, giving users the ability to program up to 48 tests. The analyzer’s flexibility features allow users to test samples from the same vial at two different temperatures, as well as test samples from either carousel in any chosen order. All models can measure samples with a viscosity range from 1 mm²/sec to 2,000 mm²/sec at 40°C.
Safe and cost-effective to own and operate, the constant pressure viscometer has a precision that meets or exceeds ASTM D445 or its equivalents.
Big Performance in a Small Package
OptiMVD measures viscosity according to ASTM D7945. This method is based upon the Hagen-Poiseuille principle of capillary flow. To determine viscosity with OptiMVD, a sample is drawn from a capped sample vial and then introduced into the measuring cell at a controlled, specified temperature. The measuring cell contains a horizontal capillary tube with optical sensors. A thermal block surrounds the measuring cell.
Density is determined by ASTM D7777 method. The OptiMVD measures density at multiple temperatures yielding precision that is much better than other D7777 density meters.
Benefits:
- Fast – quick testing at 10 mins/sample with maximum cleaning
- Precision – best-in-class with no relative bias to D445 for diesel and lubricant testing
- Flexible – sample from same vial can be used to measure viscosity at same temperature or different temperatures
- Dual viscometers/density meters models for increased sample throughput
- Optimal temperature range for most diesels/lubricants (+15°C to +100°C)
- Full viscosity range without capillary change
- Automatic Viscosity Index (VI) calculation from same vial of sample – no risk of sample contamination or mislabeling leading to inaccurate VI
- Fully automatic self-cleaning
- Small footprint saves valuable lab bench space
- Unique Smart Sampling Ordering automatically optimizes performance and decreases total test time
- Comprehensive Instrument Models: Single Sample Injection Port, 1×24 position carousel, or 2×24 position carousels
Key Features
Fast – quick testing at 10 mins/sample with maximum cleaning
Best-in-class precision for diesel and lubricant testing
Flexible viscosity testing at same temperature or different temperatures
Dual viscometers/density meters models for increased sample throughput
Full viscosity range without capillary change
Automatic Viscosity Index (VI) calculation from same vial of sample
Small footprint saves valuable lab bench space
Unique Smart Sampling Ordering for optimized performance
Specs
Test Methods & Precision
Standard Test Methods:
Viscosity : ASTM D7945
Density : ASTM D7777
Correlated Methods:
Viscosity : ASTM D445, ISO 3104, IP 71, GOST 33, GB/T 265
Density : ASTM D4052
Viscosity Repeatability:
0.5% @ 40°C
0.75% @ 100°C
Viscosity Reproducibility:
0.8% @ 40°C
1.38% @ 100°C
Density Repeatability: 0.006081 (1000 – x) kg/m3
Density Reproducibility: 0.008545 (1000 – x) kg/m3
Temperature Repeatability: ± 0.005°C
Operation
Detection:
Viscosity: Constant Pressure Capillary
Density: Oscillating U-Tube
Heating/Cooling System: Internal Peltier cooler
Test Time: Varies with required precision (10 min/sample with extensive cleaning)
Sample Loading: Automatic – Integrated sample changer (capacity for 24 samples per carousel, up to 2 carousels per analyzer)
Single sample – Single sample injection port option available
Application Range
Temperature Range: Sample: +15⁰C to +100⁰C
Viscosity Range: 1 mm2/sec to 2000 mm2/sec at 40⁰C
Density Range: 0.5 g/cm3 to 2 g/cm3
Auto Sampler Heater: Ambient to +50⁰C
Documentation
Printing: Any printer with USB Interface and PCL5 or higher, automatic print out
Result Documentation: 5000 test runs (results)
Interface
User Interface: 13.3” color touchscreen (16:9 display, 1920×1080 res.), virtual keyboard. Languages: English, German, French, Chinese and others
Interfaces: 3 X USB (2.0 full speed), 1 X Ethernet (10/100/1000 Mbps), 1 X RS-232
LIMS interface: Ethernet or RS232, automatic LIMS transfer
Data Export: USB memory stick, import into Excel
Password Protection: Different access levels for User, Admin or Service
Accessories: Barcode reader, printer
Utility Requirements
Electrical: 100V to 240VAC ±10% switchable; 50-60Hz ; 450W ; 5-2.4A
Operating Conditions: Temperature: +10°C to +30°C; recommended +15° to +25°C. Humidity: 10% to 80% rel. non-condensing
Storage temperature: +5⁰C to +55°C, 10% to 85% humidity
Dimensions: 35.6 cm (14”) wide, 49.5 cm (18.75”) deep, 47.6 cm (18.75”) tall
Weight: 43 kg (95 lb) for full load model