State-of-the-Art Solution
Product Overview
Injecting a pressurized petroleum stream into an analytical system such as a Gas Chromatograph (GC) or ElemeNtS is a challenge. Uncontrolled self-expansion of the Liquid sample can lead to discrimination effects, which impacts method precision negatively.
The ACCURA sampling device is designed for fully automated analysis of gaseous and pressurized liquid hydrocarbon streams for ElemeNtS and Gas Chromatography applications.
The ACCURA vaporizes the sample under controlled conditions, maintaining full sample integrity. ACCURA provides excellent precision and accuracy, and is extremely simple to operate. Moreover, it will tell the operator when line pressures are low enough and it is considered safe to disconnect the sample bomb from the inlet, providing unsurpassed safety.
Applications: Sulfur, Nitrogen, Nitrogen & Sulfur, Nitrogen & Halides, Sulfur & Halides, Nitrogen & Sulfur & Halides, Applicated GC Gas Analysis, Natural Gas Analysis, Refinery Gas Analysis, Sulfur Trace Analysis, Customized Analysis
Key Features
Excellent performance
Improved ease of use and more flexibility
Operator safety
Proven compliancy
Specs
Pressure
LPG and Gaseous samples with internal pressure up to 30 bars (435 psi)
Voltage
110-230 VAC
Frequency
50- 60 Hz, 240VA
Current
20 Amp
IP Class
IP21
Safety Class
Class 1
Over voltage category
CAT II
Dimensions
200mm x 480mm (7.87” x 18.9” )
Minimum Sample Pressure
2 bar (29 psi)
Methods
Accura is in line with current methods in GC and Elemental Analysis:
GC: ASTM D2163, D1945, 1946, DD2504, D2505, D2593, D2712, D4424, D5303, D6228
Elemental: D6667, D7751, UOP 936, EN17178
Software
- Operate through clear 4.3” TFT touchscreen
- All important parameters are logged at time of injection
- Temperature and pressure controls in °C/F, bar/psi/kPa.