HomeLogin / RegisterMy CartContact UsCareers

PAC Corporate Headquarters

8824 Fallbrook Drive
Houston, TX 77064 USA
Phone: 800.444.TEST
Fax: 281.580.0719

Lab InstrumentsProcess AnalyticsParts StoreCustomer SupportNews and EventsAbout PAC
Browse by:
spacer
Reformulyzer
The Reformulyzer M3 is the globally accepted industry standard for gasoline and gasoline blend products testing.
Reformulyzer

Reformulyzer

Reformulyzer

Hydrocarbon Group Type Analysis

Test Methods
ASTM D5443, ASTM D6839, EN 14517, IP 382, IP 566, ISO 22854, EN ISO 22854

The Reformulyzer® M3 is the industry standard for quantitative analysis of saturates, olefins, aromatics, and oxygenates in spark-ignition engine fuels by multidimensional gas chromatography. The Reformulyzer is based on a multi-dimensional column switching technique. After sample introduction, a set of columns and traps separates the sample into different hydrocarbon groups. The multi-dimensional technique determines carbon number distribution for the different component classes.

Operation is simple with its user-friendly software interface. Powerful, integrated diagnostic tools enable automated validation through quality control samples.

The application resides on an easy-to-use, compact Agilent Technologies 7890A gas chromatograph, configured with

  • A flame ionization detector (FID)
  • TPI inlet with electronic pneumatics control (EPC)
  • All valves, columns, and traps are mounted in a separate extension of the system for easy access

Applications

Multidimensional Group Type, Aromatics, Olefin, Oxygenates, Gasoline Characterization, Gasoline

Features & Benefits

High throughput

  • Analyzes within 75 minutes

Enhanced features

  • Offers total gasoline analyses in one injection
  • Features preset analyzer modes for FCC, straight run naphtha, alkylates, isomerates, reformates, jet fuel, and oxygenated gasolines
  • Includes an extensive suite of preset methods, sample types, and calibration/QC tools
  • Generates comprehensive group-type reporting with required specific calculations
  • Offers an analysis simulation program for diagnostic/training purposes
  • Uses dedicated connections for a leak-tight system and simplified part replacement
  • Features automated conditioning capabilities for columns and traps
  • Provides excellent LIMS connectivity; upload the results directly to your LIMS or manually after approval
  • Ensures low maintenance & easy access through a side cabinet

Fully automated

  • Offers easy validation through dedicated quality control samples and integrated diagnostic tools
  • Features optional Automatic Liquid Sampler for automation

Meets standards

  • Complies with ASTM D5443, D6839, IP 382, IP526, IP566, EN14517, ISO 22854, and the European gasoline specifications (EN 228), which includes the EN ISO 22854 (multidimensional GC method with Reformulyzer) as a referee method for hydrocarbon content, olefins and aromatics, benzene, oxygenates, and oxygen content

Brochure